Aerodynamic particle size tests in aerosol inhalation toxicology studies.
Aerodynamic particle size tests of particles in environment.
Aerodynamic particle size tests of dust in occupational health workplaces.
It is a low-flow particle size analyzer by adopting inertial impact principle, used for stage sampling and weighing of aerosol particles and then calculating parameters such as MMAD and GSD, etc.
Sampling flow rate of 0.5L/min is ideal for particle size analysis in animal exposure tests.
Sampler is divided into 7 stages + membrane stage, cutting particle size of each stage as below:
Stage |
Cutting Diameter (μm) |
1 |
5.1 |
2 |
3.1 |
3 |
1.9 |
4 |
1.2 |
5 |
0.77 |
6 |
0.50 |
7 |
0.33 |