Aerodynamic particle size tests in aerosol inhalation toxicology studies.
Aerodynamic particle size tests of particles in environment.
Aerodynamic particle size tests of dust in occupational health workplaces.
Model HRH-APS05 aerodynamic particle sizer is a low-flow particle size analyzer by adopting inertial impact principle, used for stage sampling and gravimetric filter analysis and then calculating MMAD and GSD, etc.
Sampling flow rate of 0.5L/min is ideal for particle size analysis in animal exposure tests.
Sampler is divided into 7 stages + membrane stage, cutting particle size of each stage as below:
Stage |
Cutting Diameter (μm) |
1 |
5.1 |
2 |
3.1 |
3 |
1.9 |
4 |
1.2 |
5 |
0.77 |
6 |
0.50 |
7 |
0.33 |