Designed by Mie Scattering principle, HRH-APM283 aerosol photometer is mainly used for integrity testing of high-efficiency filter on site. 7-inch LCD screen can display parameters of concentration data, leakage rate in real-time. Suitable for HEPA filters, HVAC system, biosafety cabinets, clean rooms, glove boxes and other leakage detection.
Long-life LED light.
High precision photomultiplier detection.
Support DOP, PAO and other aerosol media.
7-inch high-definition LCD screen.
Portable scanning probe can display all host status.
Large capacity data storage, real-time preservation of adopted data.
Sound and light alarm function.
Real-time printing of test data.
Export or print historical data via USB flash disk.
Dimensions |
355X265X252mm |
Weight |
10kg (host), 0.5kg (probe) |
Power supply |
220VAC 50/60HZ |
Display range |
0.0001%~100.0% |
Dynamic range |
0.0001~600ug/L |
Alarm |
built-in sound and light, control panel and scanning probe will alarm when detected concentration exceeds set point. Sound alarm function can be canceled at any time |
Settable aerosols |
PAO, DOP, etc. |
Accuracy |
1% of reading value (0.01%~100%) |
Repeatability |
0.5% of reading value (0.01%~100%) |
Light source |
long life solid LED |
Flow control |
28.3 l/min ±5% |
Auto zero |
establish zero point automatically when startup |
Report mode |
continuous mode, storage mode, print mode |
GB 50591-2010 Code for Construction and Acceptance of Cleanroom
GB 50073-2013 Code for Design of Clean Room
YY 0569-2011 Class II Biological Safety Cabinets
JJF 1800-2020 Calibration Specification of Aerosol Photometers
JJF 1815-2020 Calibration Specification for Class II Biosafety Cabinets
ISO 14644-3:2005 Cleanrooms and associated controlled environments- Part 3: Test methods
Good Manufacturing Practice for Drugs (2010 Revision)- Premises and Facilities